NSLS BEAMLINE DESCRIPTION: U10B - Infrared


GENERAL INFORMATION

PARTICIPATING INSTITUTIONS
BNL-NSLS, Northrop Grumman Advanced Systems and Technology
LOCAL CONTACT / BEAMLINE MANAGER: Larry Carr
(516) 344-2237, carr@bnl.gov
SPOKESPERSON: Larry Carr
(516) 344-2237, carr@bnl.gov

INSTRUMENT


OPTICAL CONFIGURATION

Beamline Optical System

A two-mirror system (M1 and M2) collects and re-images the synchrotron infrared source at a point just outside of the storage ring's UHV.  M1 is a water-cooled plane mirror made from silicon with a gold reflective coating.  M2 is a glass ellipsoid with an aluminum reflective coating.  The ellipsoidal mirror focuses the beam through an 11mm aperture wedged diamond window (~350 microns thick).  Delivered spectral range extends from approximately 10 cm-1 to beyond 40,000 cm-1. The infrared is then collimated to a diameter of 14mm or 8mm and transported under rough vacuum through a KBr window and into the Irus endstation.

Experimental Apparatus

Scanning infrared microspectrometer, consisting of a compact FTIR system with fixed Ge/KBr beamsplitter, integrated to an IR microscope. Includes 15X (N.A. = 0.58) and 32X (N.A. = 0.65) Schwarzchild all-reflecting objectives.  Working distance up to ~1cm. Two glass objectives (4X and 20X plan) for visual inspection.  Automated X-Y scanning stage for spectroscopic mapping.  Step resolution of 1 micron.  Small area MCT, photoconductive Ge:Cu detectors.

Computer System Hardware and Software:

Intel Pentium based PC running Spectra-Tech / Nicolet Atlus and Omnic software under Windows 95.  Includes video image capture of sample specimen.  HP LaserJet (B&W) and Deskjet (color) printers for hardcopy output.  Software for converting data into simple (ascii) X,Y data format.  Both 1.44Mb diskette and 100Mb Zip disk drives.  TCP/IP networking with Telnet, Netscape, and FTP (both client and server).  Large data sets can be archived to CD-ROM or Jazz (1Gb) media.

U10 IR Microspectrometer Beamline


For information please contact G.L. Carr
DISCLAIMERS : Revised Date : 16-Aug-98